INJECTOMAT AGILIA, THE NEW GENERATION OF STAND ALONE SYRINGE PUMPS
• flow rate range 0.1 - 1200 ml/h in normal mode, depending on the syringe capacity (0.1 ml/h increment)
•flow rate accuracy ± 1% on mechanism ; ± 2 % on syringes. syringe capacities 5, 10, 20, 30/35, 50/60 CC
• Types of syringe upto 50 types infusion modes : ml/h (0.1 ml/h increments). Volume/time: Volume to infuse 0.1 to 99.9 ml/Time: 0 h 01 to 96 h 00
• Volume limit: 0.1-999.9 ml (0.1 ml increments).priming 3 modes: mandatory, not mandatory, or advised/Rate: 1200 ml/h
• Bolus Manual bolus: Rate: 50 - 1200 ml/h (50 ml/h increment). end infusion (v/t & vl ) KVO 1 and KVO 2: from 0.1 to 5 ml/h adjustable and storable, continuous infusion or stop. fast start By default.
• Pause Programmable from 1 minute to 24 hours, increments from minute to minute.
• Night mode decreases the brightness of the screen and the green lights. The key beep can eventually be turned off. The night mode can be programmed manually or automatically in a variable time range.
• Drug libraries to be created with Vigilant, the IV Medication Safety Solution. • Adjustment of the dilution, authorised infusion modes, default infusion modes, flow rate and bolus limits in ml/h. upto 50 drugs.
Pressure management:
• Pressure modes 2 modes available: variable or 3 pre-set levels - from 100 to 900 mmHg, 50 mmHg increment.
• DPS: The Dynamic Pressure System - warns of pressure variations
• A risk of obstruction or a possible leak in the infusion line can thus be anticipated
• Pressure monitoring Graphic representation of the pressure limit thanks to the pictogram
•Anti-bolus system Reduces significantly bolus after occlusion release (0.2 ml max.).
Availability date:
50 years of experience
Large distribution network
Quality technical services
More info
Alarms-pre-alarms-security:
• Pump status Green for infusion in progress, orange for pre-alarm and warnings, red for alarm. Visibility upto 5 meters.
•All alarms are expressed by means of light indicators, written words, pictograms and sound beeps
•Syringe position control Syringe barrel clasp check, plunger head detection, anti-siphon system check, flange detection
• Infusion control Occlusion pressure pre-alarm, occlusion pressure alarm, end of infusion pre-alarm, end of infusion alarm
• Volume limit pre-alarm, volume limit alarm, keypad locking, hard and soft flow rate limits
Device control:
• Disengaged driving mechanism alarm,
• Low battery pre-alarm,
• Discharged battery alarm,
• Battery capacity display in hours and minutes,
• Unconfirmed programming,
• Technical malfunction alarm (auto-test, rotation),
• Drive system advance check,
• Watchdog check,
• Communication connection failure
• Maintenance Preventive warning.
Technical specifications:
• Manual pusher Protection for the ongoing infusion thanks to “Push-Guard”
• Display Blue graphic LCD monochrome, size 66 mm x 33 mm (128 x 64 pixels)
• Swinglock clamp Versatile clamp that allows the fixation on a rail or on a pole (Pole: 20-40 mm max. / Rail: 25-35 x 10 mm)
• Stackability upto 3 devices self-stackable on a pole with a safe handle
• Dimensions (h/w/d) / weight 160 x 345 x 135 mm / 2150 g
•Battery LIFE NI/MH. Capacity minimum: 10 hours at 5 ml/h.
• Battery replacement: separate compartment – no need for electrical security testing
• Power suply 100 - 240 Vac / 15 VA, external power 9 Vdc by a converter.
2 other products in the same category:
SYRINGE PUMP • Compatible with syringe of any standard , 5, 10, 20, 30, 50, 60 ml • Infusion limit volume range: 0.1~1000ml (in 0.1, 1, 10 ml increments) Flow Rate: • Syringe 5 ml : 0,1 - 100 ml/h• Syringe 10 ml : 0,1 - 300 ml/h•...
SYRINGE PUMP 2 CHANNELS • Compatible with syringe of any standard , 10, 20, 30, 50, 60 ml • Infusion limit volume range: 0.1~10000ml (in 0.1, 1 ml increments) Flow Rate: • Syringe 10 ml : 0,1 - 420 ml/h• Syringe 20 ml : 0,1 - 650 ml/h•...